'The Common Mode Voltage is +/-11V maximum (operational). With a 60Hz +/-11V input on both sides of the differential inputs, with 1K in series with 1 side the
Common Mode Rejection Ratio in -74dB. 22VPP input = 4.4millivolts pp on the output. This specification is important in high
power line noise applications. *Created: 12/10/03